Pfeiffer Vacuum, a leading provider of vacuum technology, introduces the new leak tester SpeedAir 3050, an all-in-one solution for nonporous pharmaceutical containers. SpeedAir offers a CCIT solution for a wide range of products: Whether the products are flexible or rigid, liquid or solid, the SpeedAir can swiftly and accurately test them all. Typically requiring only 30 – 45 seconds, this mass extraction instrument quickly and efficiently delivers results.
With testing times as short as 30 – 45 seconds, it delivers reliable results, ensuring the integrity of pharmaceutical containers. Unlike alternative methods that focus on specific areas or access points, the SpeedAir provides a comprehensive global Non-Destructive Testing (NDT) approach for containers. It adheres to ASTM Standard F3287-17, demonstrating the capability in independent third-party labs to 1 micron using an air-based technology.
Product Specifications
Test method: Mass Extraction
Power: 90–250 V AC / 50-60 Hz
Flow sensor: IMFS (Intelligent Molecular Flow Sensor)
Sensitivity to 1 micron defect size
Operating conditions:
Temperature: (min./max.) 15–45°C (59–83°F)
Humidity: (min./max.) 30–80 %
Test Pressure Range: 1–20 Torr (1.3–27 mbar abs)
Pneumatic Supply: Air (pilot operated valves)
Quality: 1.3.1 according to ISO 8573-1
Pressure: (min./max.) 4.5/10 bar rel. (65–145 psig)
Nitrogen supply Recommended for chamber venting
Pressure: (min./max.) 140/690 mbar rel. (2–10 psig)
Operating System: Windows 10
User Interface: 10″ Multi-touch Full HD color screen
LeakRx Software 21 CFR part 11 compliant
Stored Test Recipes: Unlimited
Network connection: 1 x LAN (RJ45)
Interfaces: USB, Wired Ethernet
Discrete Inputs: Start, Stop, Test Type Change
Analog Inputs: 2 x 0–5V
Discrete Outputs: Clamp, Exhaust, Custom, Test Type, A-D
Analog Outputs: 2 x AO
Dimensions (l x w x h):
Display closed 922 x 601 x 1,129 mm (36 x 24 x 44 inch)
Display open 922 x 601 x 1,397 mm (36 x 24 x 55 inch)
Weight: (Including trolley) 127 kg / 280 lbs
Noise level: < 53 dB(A)